P17080: Heart Pump and Circulatory System
/public/

Build & Test Prep

Table of Contents

Action Items Review and Status

Review of Action Items

Review of Action Items

Updated Bill of Materials and Parts Status

Detailed Parts Status

Detailed Parts Status

Updated BOM

Updated BOM

Problem Review

Problem Matrix

Problem Matrix

Problem Matrix Tracking

Problem Matrix Tracking

See Live Document Here: Problem Matrix

Design Updates

Heart Pump Updates

The drawings were updated so that the Pneumatic Slot Ring was machinable and to make final mentioning and location adjustments of holes. Two options were generated. One mimics the original shape and the other is a circle. Option 1 is more desirable because it fits closer to the base of the ventricle. Option 2 was generated in the event that machining option 1 is too difficult or not probable.

Ventricle Hydraulic Ring Update

Ventricle Hydraulic Ring Update

Ventricle Pneumatic Cap Update Option 1

Ventricle Pneumatic Cap Update Option 1

Ventricle Pneumatic Slot Ring Update Option 2

Ventricle Pneumatic Slot Ring Update Option 2

Ventricle Pneumatic Cap Update

Ventricle Pneumatic Cap Update

See Live Drawings Here: Updated CAD Drawings

Labview Updates

The flow meter program was adapted to use a counter to receive more accurate measurements of flow. This program is optimized in comparison to the original program and produces data without significant noise.

Flow Meter LabVIEW Update

Flow Meter LabVIEW Update

Build Status

The Ventricle Hydraulic Ring was machined and the holes are tapped so that the connections to the valves can be attached easily. The Ventricle Pneumatic Cap was machined and is show fitting with the acrylic cylinder. A o-ring may be used to help with the seal.

The Ventricle Pneumatic Slot Ring was not machined past drilling the outer holes. This is because the drawings were being adapted to make the slot machinable. The original problem was that the drawing was made using ellipses and the machine shop can not produce that shape without knowing radii.

Ventricle Hydraulic Ring

Ventricle Hydraulic Ring

Ventricle Pneumatic Cap

Ventricle Pneumatic Cap

 Current Build Status

Current Build Status

Testing

Water Source Testing

Bacterial film buildup testing:
 Water Source Testing

Water Source Testing

Demonstrates visual bio-film buildup using faucet water and no visual bio-film on inner lining of De-ionized water.

Due to absence of visual bio-film buildup within the inner tubing using De-ionized water, this will be used as a water source for the heart pump and circulatory system model.

Lab Cleaning Solution Testing =

Testing silicone degradation testing via Alconox solution and Isopropyl alcohol:

 Lab Cleaning Solution Testing

Lab Cleaning Solution Testing

Submerged silicone sections in Alconox solution and Isopropyl alcohol solution.

Due to absence of material property changes using Alconox solution for both types of silicone, Alconox will be used as the lab cleaning solution for either type of silicone.

Flow Meter Calibration

To determine the conversion from revolutions counted to L/min:

Calibration Set-up

Calibration Set-up

Tank Reserve Volume

Tank Reserve Volume

Water Flow

Water Flow

See Raw Data Sets Here: Raw LabVIEW Data

Overall Accomplishments Since Last Review

Schedule Review

Original Schedule

Original Schedule

Original Schedule

Updated Schedule

Updated Schedule

Updated Schedule

Review Action Items and Three Week Plans

Review Action Items

Individual Three Week Plan

Time in days is referencing the amount of work days needed to complete each task.

Susan

Kevan

Blake

Mike

Henry